ECS-1812 is for laboratory use to calibrate vibration transducers which are working during low frequency range. The essential part of ECS-1812 Is a standard shaker with very low working frequency. This standard shaker is built with air bearing guidance with extreme low friction. Working with an innovative active feed-forward control technique, ECS-1812 has excellent performances in low frequency calibration.

The calibration method of ECS-1812 is also the secondary calibration (back to back), to measure the sensitivity of the sensor-under-test (SUT) by point to point swept sine or random vibration over standard or user-defined frequency and acceleration ranges.

ECS-1812 is suitable for calibration of those sensors used for earthquake monitoring, MEMS products development, and industries of railway, oil & gas, electricity, etc. 
 

Features

  • ISO 16063-21 standard compliant
  • Calibration frequency range is 0.1-200 Hz, expandable to 0.01Hz
  • Support to calibrate all kinds of vibration accelerometers, velocity and displacement sensors
    (voltage, PE, IEPE, Piezo-resistive, Capacitive) , etc.
  • Multiple vibration excitation mode of point to point, swept sine and random
  • Loop control calibration, and optional open loop calibration available
  • Automatic procedure, quick and accurate calibration for instant results
  • Support calibration raw data export
  • Customized calibration report generation
     

System Model

ECS-1812

Max. Excitation Force

400 N

Calibration Frequency Range

0.1-200 Hz, expandable to 0.01Hz

Max. acceleration

1 g

Max. Displacement (p-p)

100 mm

Max. Velocity

0.2 m/s

Max. Load

10 kg

System Uncertainty

<1% (160Hz, 100 m/s2)

Calibration Method

secondary (back to back)

Calibration Function

sensitivity, frequency response (amplitude and phase), and amplitude linearity

Calibration Procedure

point to point, swept sine and random excitation